Up to 150 times (Tip replacement required after 150 sterilization cycles)
️⚠️Caution⚠️️
The condition of the internal mirror of the tip can greatly affect the quality of scan data. Always check the mirror before scanning.
If there are foreign substances or stains on the internal mirror, clean it thoroughly before use.
If the mirror is scratched, damaged, or has stains that cannot be removed, discard the tip and replace it with a new one.
If the tip is sterilized in an unsealed sterilization Bag (or pouch), the mirror may be damaged. To prevent this, always seal the sterilization Bag (or pouch) before sterilization.
Damaged scanner tips must be discarded. Using a damaged tip may cause unintended injury to the patient.
Scanner tips can be repurchased by contacting the manufacturer or distributor. (📩 Contact link)
For the front optical lens of the main body, as with the mirror inside the tip, care must be taken to avoid foreign substances and smudges.
Do not place the scanner in a sterilizer or immerse in water or disinfectant solutions.
Gloves must be worn when cleaning and disinfecting the scanner.
Excessive liquid may damage the scanner.
When disinfecting the scanner, do not use cotton products, clothing or tissues soaked in disinfectant.
Read and follow the warnings and personal protection instructions provided on the safety data sheet (SDS) for the disinfectant used on the scanner for reuse.