📌 PIM KIT Cleaning And Sterilization Guide
📌 PIM KIT Storage And Maintenance Guide
📌 PIM KIT Cleaning And Sterilization Guide
The PIM Scan Body is supplied in a non-sterile condition. For patient safety, it must be sterilized before use.
Cleaning and sterilization should be performed according to the instructions below.
Using a product that has not been properly sterilized may lead to infection; therefore, always ensure the scanbody is sterile before use.
1. Cleaning And Sterilization Materials
Dry Cleaning Cloth |
Diluted Detergent (Endozime) |
Soft Nylon Brush |
Sterilization Bac |
|
Detergent Concentration: 0.4% Diluted Solution |
2. Cleaning And Sterilization Method
- PIM Scanbody & Cap Scanbody
[ Cleaning Method ] |
|
① Rinse under running water and gently remove contaminants using a soft brush. ② Soak in a pH-neutral, enzymatic detergent for 5–10 minutes, following the detergent manufacturer's instructions. (Do not soak the device in enzymatic detergent for longer than recommended.) ③ Rinse thoroughly under running water to remove any remaining detergent. ④ Dry completely before packaging. (For cloth, use a dry, lint-free cloth) ⑤ Place the device in appropriate sterilization packaging. ⑥ Autoclave according to the recommended sterilization conditions. |
️⚠️Caution⚠️️
|
- PIM Guide Tool and Cover
[ Cleaning Method ] |
|
① Wipe off contaminants with running water for more than 30 seconds. ② Soak in diluted detergent (Endozime) for more than 30 seconds. ③ Wash with flowing purified water for more than 30 seconds. ④ Remove moisture by wiping with a clean non-absorbent or soft cloth. ⑤ If the stain is still present after following the above procedure, replace the guide and cover. ⑥ Wiping with strong force may scratch the lens, which may affect the quality of the scanned data. ⑦ Place cleaned, dried guide and cover in a sterile bag (or pouch) and seal tightly. ⑧ Place the guide and cover in the autoclave while the bag (or pouch) is still sealed. |
3. Steam (Autoclave) Sterilization Conditions
Autoclave Sterilization Conditions | |
134℃ / 273.2℉ |
Heating for 15 minutes / Drying for30 minutes |
⚠️Do not use non-recommended sterilization methods: | |
️⚠️Caution⚠️️
|
📌 PIM KIT Storage And Maintenance Guide
PIM KIT Storage Method |
|
PIM KIT Maintenance Method |
|